TERRY-SCHANZLIN ASTIGMATOME 04905

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2007-10-18 for TERRY-SCHANZLIN ASTIGMATOME 04905 manufactured by Oasis Medical.

Event Text Entries

[689677] A patient who, according to the astigmatome nomagram, required two 45 degree arcuate incisions using the 8 mm optical zone. Pre-op corneal topography showed central pachymetry of about 550 microns and mid-peripheral corneal thicknesses of 580+ microns, so a 550 micron double blade was chosen. The astigmatome was employed in the usual fashion at the beginning of the case prior to the performance of the phako/iol portion of the procedure using a rezoom lens, and resulting in two full-thickness 45 degree incisions. Each incision required suturing with a figure-eight 10-0 nylon suture.
Patient Sequence No: 1, Text Type: D, B5


[8011077] Oasis medical is communicating with the customer to have the customer return the product in question, and oasis medical will evaluate and submit a evaluation summary when it is available. As of the date of this initial mdr, the product has not veen received. Various communications with the customer have not been able to obtain the lot number or detailed description of follow-up events.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2083373-2007-00006
MDR Report Key929670
Report Source04
Date Received2007-10-18
Date of Report2007-10-18
Date Mfgr Received2007-05-16
Date Added to Maude2007-12-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactROBYN SCOPIS, MANAGER
Manufacturer Street514 S. VERMONT AVE.
Manufacturer CityGLENDORA CA 91741
Manufacturer Postal91741
Manufacturer Phone6268525156
Single Use3
Remedial ActionPM
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTERRY-SCHANZLIN ASTIGMATOME
Generic NameASTIGMATOME INSTRUMENT SET
Product CodeHNC
Date Received2007-10-18
Model Number04905
Catalog Number04905
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeNA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key924707
ManufacturerOASIS MEDICAL
Manufacturer Address514 S. VERMONT AVE. GLENDORA CA 91741


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2007-10-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.