C8I ? OCT INTEGRATED SYSTEM C408652

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-11-15 for C8I ? OCT INTEGRATED SYSTEM C408652 manufactured by St. Jude Medical.

Event Text Entries

[166571650] The results/method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10


[166571651] After a power surge in the hospital, an audible "bang" occurred and smoke was noted to be coming from the system. There was intermittent angio co-registration feed resulting in grainy images being displayed. The hospital experienced a power surge and the system became unable to function properly. The fuses were replaced and the kit was attempted to be booted, but an audible "bang" occurred and the system would no longer power up. Smoke was noted to be coming from the system. There was no patient involved in this event. After investigation, the power supply unit fuses were found to be blown. The fuses were replaced and the issue was resolved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009600098-2019-00027
MDR Report Key9328659
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-11-15
Date of Report2019-11-15
Date of Event2019-10-31
Date Mfgr Received2019-10-31
Device Manufacturer Date2018-11-27
Date Added to Maude2019-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL
Manufacturer Street4 ROBBINS DRIVE
Manufacturer CityWESTFORD MA 01886
Manufacturer CountryUS
Manufacturer Postal Code01886
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameC8I ? OCT INTEGRATED SYSTEM
Generic NameSYSTEM, IMAGING, OPTICAL COHERENCE TOMOGRAPHY (OCT)
Product CodeNQQ
Date Received2019-11-15
Model NumberC408652
Lot Number6802647
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL
Manufacturer Address4 ROBBINS DRIVE WESTFORD MA 01886 US 01886


Patients

Patient NumberTreatmentOutcomeDate
10 2019-11-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.