ANAEROBIC PULSATORT PLUS ARTERIAL BLOOD SAMPLING KIT 4058NP-2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2007-10-30 for ANAEROBIC PULSATORT PLUS ARTERIAL BLOOD SAMPLING KIT 4058NP-2 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[709041] User alleges two events of blood spray into the eyes due to snapping the needle into the needle protection, causing the blood to spray. No treatment reported.
Patient Sequence No: 1, Text Type: D, B5


[7882803] Results evaluations codes: our investigation into this event is very limited as the user facility did not retain the needle used in this event. A review of our complaints database show no other reports on this device lot number of the finished good or the needle lot number. A review of the labeling reveals this issue is identified in our instructions for use: a review of the instructions for use reveals: and a work step "after procedure is completed, press the needle into the sheath using a one-handed technique by gently pressing the sheath against a flat surface. " no corrective action has been assigned as this event is likely due to user error. This report has been logged for trending purposes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217052-2007-00137
MDR Report Key934916
Report Source07
Date Received2007-10-30
Date of Report2007-10-02
Date Mfgr Received2007-10-02
Date Added to Maude2007-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTIMOTHY MASON
Manufacturer CityKEENE NH 03431
Manufacturer Postal03431
Manufacturer Phone6033523812
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer CityKEENE NH 03431
Manufacturer Postal Code03431
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANAEROBIC PULSATORT PLUS ARTERIAL BLOOD SAMPLING KIT
Generic Name75 GJE - TRAY, BLOOD COLLECTION
Product CodeGJE
Date Received2007-10-30
Catalog Number4058NP-2
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key932766
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address10 BOWMAN DR. KEENE NH 03431


Patients

Patient NumberTreatmentOutcomeDate
10 2007-10-30

© 2026 FDA.report