Event Text Entries

[168050324] The manufacturer has received the sample and will evaluate. Results are expected soon. A lot history review (lhr) of recv0302 showed no other similar product complaint(s) from this lot number.
Patient Sequence No: 1, Text Type: N, H10


[168050325] It was reported that leakage from the catheter was found while using the power trialysis. There was no reported patient injury. On (b)(6) 2019 the facility reported that the initial information was incorrect and the correct event description was as below. (b)(6). It was reported that during insertion of the catheter, the catheter could not be advanced/withdrawn because the guidewire was stuck in the catheter. There was no reported patient injury. On 11/04/2019 - returned wire is unraveled.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006260740-2019-03666
MDR Report Key9370933
Report SourceHEALTH PROFESSIONAL
Date Received2019-11-25
Date of Report2019-11-25
Date of Event2019-10-02
Date Mfgr Received2019-11-04
Device Manufacturer Date2018-08-01
Date Added to Maude2019-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKELSEY ERICKSON
Manufacturer Street605 N. 5600 W.
Manufacturer CitySALT LAKE CITY UT 84116
Manufacturer CountryUS
Manufacturer Postal84116
Manufacturer Phone8015225937
Manufacturer G1BARD REYNOSA S.A. DE C.V. -9617592
Manufacturer StreetBLVD. MONTEBELLO #1 PARQUE INDUSTRIAL COLONIAL
Manufacturer CityREYNOSA, TAMAULIPAS
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOWER-TRIALYSIS SLIM-CATH STRAIGHT SHORT-TERM DIALYSIS CATHETER (12F) (TRIPLE-LU
Generic NameCATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Product CodeNIE
Date Received2019-11-25
Returned To Mfg2019-10-23
Model NumberN/A
Catalog Number5853200J
Lot NumberRECV0302
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBARD ACCESS SYSTEMS
Manufacturer Address605 N. 5600 W. SALT LAKE CITY UT 84116 US 84116


Patients

Patient NumberTreatmentOutcomeDate
10 2019-11-25