MTS INCUBATOR MTS9680

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2007-11-13 for MTS INCUBATOR MTS9680 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[728923] The customer indicated that the incubator generated smoke. The customer had a 3 gallon container of saline above the incubator. The saline leaked inside the incubator and melted the circuit board inside causing smoke to generate from the inside. No fire or shock reported as a result of this incident. No death or serious injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[7902389] User error identified as root cause. Saline leaked inside incubator. Product labeling states "do not pour liquid over the instrument. " (mts incubator user's guide 2004-06-30 p. 9) assistance provided to the customer over the phone has resolved the issue. Instrument has been removed from service and disposed by customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2007-00313
MDR Report Key942952
Report Source05,06
Date Received2007-11-13
Date of Report2007-11-13
Date of Event2007-10-19
Date Mfgr Received2007-10-19
Device Manufacturer Date2002-06-01
Date Added to Maude2008-01-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactAWILDA GERBER, ASSOC III
Manufacturer Street1295 S.W. 29TH AVE.
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239567
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS INCUBATOR
Generic NameMTS GEL CARD INCUBATOR
Product CodeJTQ
Date Received2007-11-13
Model NumberNA
Catalog NumberMTS9680
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key938044
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-11-13