MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-12-12 for STERRAD? 100S STERILIZER 10101 manufactured by Advanced Sterilization Products.
[182558287]
A field service engineer was dispatched to the customer site. The catalytic converter was replaced to resolve the smoke/haze issue. Unit meets specifications and was returned to service. (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[182558288]
A customer reported an event of a "steam" or a smoke/haze emitting from the sterrad? 100s sterilizer. There was no report of any injuries or human reactions. It is unclear whether or not there was a cycle cancellation with this event and asp will continue to follow-up for further information. An asp field service engineer was dispatched to assess the unit onsite. This event is being reported as a malfunction report subsequent to a serious injury.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2084725-2019-01005 |
MDR Report Key | 9461201 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2019-12-12 |
Date of Report | 2019-11-22 |
Date of Event | 2019-11-22 |
Date Mfgr Received | 2020-02-12 |
Device Manufacturer Date | 1998-11-23 |
Date Added to Maude | 2019-12-12 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. GABRIELA MCLELLAN |
Manufacturer Street | 33 TECHNOLOGY DRIVE |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal | 92618 |
Manufacturer Phone | 949453-639 |
Manufacturer G1 | ADVANCED STERILIZATION PRODUCTS |
Manufacturer Street | 33 TECHNOLOGY DRIVE |
Manufacturer City | IRVINE CA 92618 |
Manufacturer Country | US |
Manufacturer Postal Code | 92618 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | STERRAD? 100S STERILIZER |
Generic Name | STERRAD EQUIPMENT |
Product Code | MLR |
Date Received | 2019-12-12 |
Model Number | 10101 |
Catalog Number | 10101 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ADVANCED STERILIZATION PRODUCTS |
Manufacturer Address | 33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-12-12 |