MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-12-12 for STERRAD? 100S STERILIZER 10101 manufactured by Advanced Sterilization Products.
        [182558287]
A field service engineer was dispatched to the customer site. The catalytic converter was replaced to resolve the smoke/haze issue. Unit meets specifications and was returned to service. (b)(4).
 Patient Sequence No: 1, Text Type: N, H10
        [182558288]
A customer reported an event of a "steam" or a smoke/haze emitting from the sterrad? 100s sterilizer. There was no report of any injuries or human reactions. It is unclear whether or not there was a cycle cancellation with this event and asp will continue to follow-up for further information. An asp field service engineer was dispatched to assess the unit onsite. This event is being reported as a malfunction report subsequent to a serious injury.
 Patient Sequence No: 1, Text Type: D, B5
| Report Number | 2084725-2019-01005 | 
| MDR Report Key | 9461201 | 
| Report Source | COMPANY REPRESENTATIVE,FOREIG | 
| Date Received | 2019-12-12 | 
| Date of Report | 2019-11-22 | 
| Date of Event | 2019-11-22 | 
| Date Mfgr Received | 2020-02-12 | 
| Device Manufacturer Date | 1998-11-23 | 
| Date Added to Maude | 2019-12-12 | 
| Event Key | 0 | 
| Report Source Code | Manufacturer report | 
| Manufacturer Link | Y | 
| Number of Patients in Event | 0 | 
| Adverse Event Flag | 3 | 
| Product Problem Flag | 3 | 
| Reprocessed and Reused Flag | 3 | 
| Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL | 
| Health Professional | 3 | 
| Initial Report to FDA | 3 | 
| Report to FDA | 3 | 
| Event Location | 3 | 
| Manufacturer Contact | MS. GABRIELA MCLELLAN | 
| Manufacturer Street | 33 TECHNOLOGY DRIVE | 
| Manufacturer City | IRVINE CA 92618 | 
| Manufacturer Country | US | 
| Manufacturer Postal | 92618 | 
| Manufacturer Phone | 949453-639 | 
| Manufacturer G1 | ADVANCED STERILIZATION PRODUCTS | 
| Manufacturer Street | 33 TECHNOLOGY DRIVE | 
| Manufacturer City | IRVINE CA 92618 | 
| Manufacturer Country | US | 
| Manufacturer Postal Code | 92618 | 
| Single Use | 3 | 
| Previous Use Code | 3 | 
| Event Type | 3 | 
| Type of Report | 3 | 
| Brand Name | STERRAD? 100S STERILIZER | 
| Generic Name | STERRAD EQUIPMENT | 
| Product Code | MLR | 
| Date Received | 2019-12-12 | 
| Model Number | 10101 | 
| Catalog Number | 10101 | 
| Operator | HEALTH PROFESSIONAL | 
| Device Availability | N | 
| Device Age | DA | 
| Device Eval'ed by Mfgr | Y | 
| Device Sequence No | 1 | 
| Device Event Key | 0 | 
| Manufacturer | ADVANCED STERILIZATION PRODUCTS | 
| Manufacturer Address | 33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618 | 
| Patient Number | Treatment | Outcome | Date | 
|---|---|---|---|
| 1 | 0 | 2019-12-12 |