DEFIBRILLATOR LIFE PAK 8

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-07-20 for DEFIBRILLATOR LIFE PAK 8 manufactured by Physio Control.

Event Text Entries

[5203] Pt. Undergoing eps study. V-tach induced. Life pak 8 defibrillator charged at 200 joules to abort v-tach but would not discharge. Attempt made at 360 joules with same results. Eps was utilizing defib. "hands off" cassette with use of r2 pads. Biomed was able to reproduce problem at 5 joules. Lead select was then hit and put on paddles and then able to discharge. Sync light was not activated during discharge. Subsequent testing had revealed unit able to discharge. Invalid data - regarding single use labeling of device. Patient medical status prior to event: invalid data. Invalid data - regarding multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. Invalid data - regarding whether event presents imminent hazard. Invalid data - whether device used as labeled/intended. Invalid data - regarding evaluation by user after event. Method of evaluation: invalid data. Results of evaluation: invalid data. Conclusion: invalid data. Certainty of device as cause of or contributor to event: invalid data. Corrective actions: no data. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9483
MDR Report Key9483
Date Received1994-07-20
Date of Report1994-07-14
Report Date1994-07-14
Date Added to Maude1994-12-13
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameDEFIBRILLATOR
Product CodeDRK
Date Received1994-07-20
Model NumberLIFE PAK 8
Device Availability*
Implant Flag*
Device Sequence No1
Device Event Key9136
ManufacturerPHYSIO CONTROL


Patients

Patient NumberTreatmentOutcomeDate
101. No Informationnvalid Deathata 1994-07-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.