MARUS MAXSTAR NARROW BACK HYDRAULIC CHAIR DC1690-115

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2007-11-16 for MARUS MAXSTAR NARROW BACK HYDRAULIC CHAIR DC1690-115 manufactured by Dental Equipment, Llc (dba Newberg Manufacturing).

Event Text Entries

[750918] It was reported that both the left and right back rest bolts on a dc 1690 dental chair backed out at the same time causing the back rest to fall down towards the floor with a patient in the chair, however, the dental assistant happened to be behind the chair at this time and caught the chair back before it fell back down towards the floor. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[8009702] According to the dental office manager, the dealer technician replaced both the left and right bolts on the chair back assembly. The chair is not operating within factory specifications. We tried contacting the distributor on numerous occasions, including mailing a certified letter, however, the distributor never called us back. We wanted to get a better understanding of how both bolts could have simultaneously backed out at the same time. The chair has already been repaired so we are unable to determine the cause of failure since we don't have the bolts he replaced. This is the first time we have become aware of a complaint of this nature.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004517290-2007-00004
MDR Report Key949587
Report Source08
Date Received2007-11-16
Date of Report2007-10-25
Date of Event2007-10-24
Date Mfgr Received2007-10-25
Device Manufacturer Date2004-10-01
Date Added to Maude2008-01-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY, MGR
Manufacturer Street11727 FRUEHAUF DR
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Manufacturer G1DENTAL EQUIPMENT, LLC (DBA NEWBERG MANUFACTURING)
Manufacturer Street705 S. SPRINGBROOK RD. SUITE B200
Manufacturer CityNEWBERG OR 97132
Manufacturer CountryUS
Manufacturer Postal Code97132
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMARUS MAXSTAR NARROW BACK HYDRAULIC CHAIR
Generic NameDENTAL CHAIR
Product CodeKLC
Date Received2007-11-16
Model NumberDC1690-115
Catalog NumberDC1690-115
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key948908
ManufacturerDENTAL EQUIPMENT, LLC (DBA NEWBERG MANUFACTURING)
Manufacturer Address705 S. SPRINGBROOK RD. SUITE B200 NEWBERG OR 97132 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-11-16