TYPE-S HEAD EXTENSION 20-CFHN-SUB2-V

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2007-04-20 for TYPE-S HEAD EXTENSION 20-CFHN-SUB2-V manufactured by Civco Medical Solutions, Med-tec Inc..

Event Text Entries

[16639789] Civco was notified that a type-s head extension joint broke near the locking mechanism and allowed the patient to fall. The patient was lying flat on their back as they adjusted themslef of the type-s head extension. The type-s head extension broke and the patient fell about 2-3 feet. The patient was evaluated by x-ray and no injury was identified.
Patient Sequence No: 1, Text Type: D, B5


[16653706] Clinician/hospital glued joint together prior to returning for evaluation which made evaluation of the original joint adhesive impossible. Patient repositioned self putting excessive pressure on worn device. Instructions for use describes and diagrams correct repositioning process of patient/clinician. Instructions for use indicate to inspect for damage/wear prior to use and not to use a damaged device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932738-2007-00001
MDR Report Key950580
Report Source06
Date Received2007-04-20
Date of Report2007-04-11
Date of Event2007-03-26
Date Mfgr Received2007-03-26
Device Manufacturer Date2003-06-01
Date Added to Maude2007-11-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street1401 8TH ST SE
Manufacturer CityORANGE CITY IA 51041
Manufacturer CountryUS
Manufacturer Postal51041
Manufacturer Phone7127278688
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTYPE-S HEAD EXTENSION
Generic NameCARBON FIBER COUCH INSERT
Product CodeIWY
Date Received2007-04-20
Model Number20-CFHN-SUB2-V
Catalog Number20-CFHN-SUB2-V
Lot Number2901
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key921898
ManufacturerCIVCO MEDICAL SOLUTIONS, MED-TEC INC.
Manufacturer Address1401 8TH ST. SE ORANGE CITY IA 51041 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-04-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.