[188933247] There has been a total of (b)(6) sold of this product code since 2013 with 3 other complaints recorded for similar occurrences. This equates to a (b)(6) complaint rate which is categorized as low. The product has not been received for evaluation. A follow-up report will be submitted once new information is acquired.
Patient Sequence No: 1, Text Type: N, H10


[188933248] At the conclusion of the procedure, when placing the sternal wires, the "insert" from the sternal wire needle driver became dislodged from the instrument and was retained in the patient's chest. Per x-ray, the patient underwent an additional surgery to partially reopen the chest wall to revolve the piece. As a result, the patient remained in the hospital for two additional inpatient admission days.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007208013-2019-00039
MDR Report Key9511537
Report SourceUSER FACILITY
Date Received2019-12-23
Date of Report2019-12-23
Date of Event2019-08-01
Date Mfgr Received2020-01-02
Date Added to Maude2019-12-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBRANDI MEATH
Manufacturer Street3034 OWEN DRIVE
Manufacturer CityANTIOCH TN 37013
Manufacturer CountryUS
Manufacturer Postal37013
Manufacturer Phone6159645290
Manufacturer G1SYMMETRY SURGICAL INC
Manufacturer Street3034 OWEN DRIVE
Manufacturer CityANTIOCH TN 37013
Manufacturer CountryUS
Manufacturer Postal Code37013
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYMMETRY CLASSIC PLUS
Generic NameSTERNAL NEEDLE HOLDER
Product CodeHXK
Date Received2019-12-23
Returned To Mfg2020-01-02
Catalog Number36-2025
Lot Number31018167
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYMMETRY SURGICAL INC
Manufacturer Address3034 OWEN DRIVE ANTIOCH TN 37013 US 37013


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-12-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.