REF BALL JNT SCREWDRIVER SHAFT 71362295

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-12-24 for REF BALL JNT SCREWDRIVER SHAFT 71362295 manufactured by Smith & Nephew, Inc..

Event Text Entries

[172580451] It was reported that during surgery, assembly pin fell out of the joint of the screwdriver tip. Parts fell inside the patient. All parts were recovered and no harm reported. Delay of less than 30 minutes. Procedure was concluded with back device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1020279-2019-04577
MDR Report Key9516207
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-12-24
Date of Report2020-03-12
Date of Event2019-12-06
Date Mfgr Received2019-12-06
Device Manufacturer Date2012-04-12
Date Added to Maude2019-12-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. SARAH FREESTONE
Manufacturer Street1450 BROOKS ROAD
Manufacturer CityMEMPHIS TN 38116
Manufacturer CountryUS
Manufacturer Postal38116
Manufacturer Phone0447940038
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street1450 BROOKS ROAD
Manufacturer CityMEMPHIS TN 38116
Manufacturer CountryUS
Manufacturer Postal Code38116
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameREF BALL JNT SCREWDRIVER SHAFT
Generic NameINSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Product CodeMDM
Date Received2019-12-24
Model Number71362295
Catalog Number71362295
Lot Number12DM06932
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address1450 BROOKS ROAD MEMPHIS TN 38116 US 38116


Patients

Patient NumberTreatmentOutcomeDate
10 2019-12-24