STERRAD? NX STERILIZER 10033

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-12-26 for STERRAD? NX STERILIZER 10033 manufactured by Advanced Sterilization Products.

Event Text Entries

[185490860] A field service engineer was dispatched to the customer site. However, details of the resolution are pending and asp will continue to follow up for the service resolution. Asp complaint ref #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[185490861] A customer reported an event of smoke/haze and odor/smell emitting from the sterrad? Nx sterilizer. There was no report of any injuries or human reactions. The customer was advised to power down the unit, discontinue its use, and vacate the room. An asp field service engineer was dispatched to assess the unit onsite. This event is being reported as a malfunction report subsequent to a serious injury.
Patient Sequence No: 1, Text Type: D, B5


[188891750] A field service engineer was dispatched to the customer site and found no problem with the unit, no odor or haze was observed. He completed successful tests and confirmed that the unit met specifications. This file is no longer a reportable malfunction since no odor or haze could be confirmed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2084725-2019-01027
MDR Report Key9521997
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-12-26
Date of Report2019-12-04
Date of Event2019-12-04
Date Mfgr Received2020-02-26
Date Added to Maude2019-12-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. GABRIELA MCLELLAN
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone949453-639
Manufacturer G1ADVANCED STERILIZATION PRODUCTS
Manufacturer Street33 TECHNOLOGY DRIVE
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction NumberZ-0744-0746-2014
Event Type3
Type of Report3

Device Details

Brand NameSTERRAD? NX STERILIZER
Generic NameSTERRAD EQUIPMENT
Product CodeMLR
Date Received2019-12-26
Model Number10033
Catalog Number10033
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerADVANCED STERILIZATION PRODUCTS
Manufacturer Address33 TECHNOLOGY DRIVE IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2019-12-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.