CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-02 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[175553796] Product analysis: analysis confirmed that when creating a portable document format (pdf) in the application and an attempt was made to send it to the universal serial bus (usb), by using "send to", the screen froze. According to the logs, an exception occurred while reopening the socket when the application came to the foreground. This is a known background foreground issue what was fixed with a new version. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175607398] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175607399] It was further reported that the programmers screen had froze when saving the pdf to the usb in the mobile programmer application. After that, the application was rebooted, but the same event occurred twice. The pdf could be saved the third time.
Patient Sequence No: 1, Text Type: D, B5


[175653057] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175653058] It was reported that there was incorrect programming/programming difficulty with the programmer. It was noted that the programmer screen froze by storing portable document format (pdf) in the universal serial bus (usb) memory stick when the mobile programmer application was run. The issue was attempted to be resolved by shutting down and rebooting the application, however the same event occurred twice. The issue was noted to resolve with a reboot. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004593495-2020-00004
MDR Report Key9538674
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-02
Date of Report2020-01-17
Date of Event2019-12-11
Date Mfgr Received2020-01-15
Date Added to Maude2020-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-01-02
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-02

© 2024 FDA.report