IFUSE IMPLANT SYSTEM 7065M-100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2020-01-02 for IFUSE IMPLANT SYSTEM 7065M-100 manufactured by Si-bone, Inc..

Event Text Entries

[173446633] Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification. The most probable root cause is user error; improper implant size selection, using an implant that was too long or placing the implant too deep.
Patient Sequence No: 1, Text Type: N, H10


[173446634] The patient had right side si joint arthrodesis in (b)(6) 2019 where three implants were installed. The patient complained of radicular pain following the procedure. The surgeon determined that there was a bone fragment caused by the installation of the superior positioned implant that may have been compressing the nerve in the neuroforamen in the s1 position. In (b)(6) 2019, a different surgeon removed the superior positioned implant and placed bone graft in the explant void. No other preexisting implants were adjusted or removed. The patient's radicular pain symptoms resolved following the revision procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007700286-2019-00155
MDR Report Key9539251
Report SourceCOMPANY REPRESENTATIVE
Date Received2020-01-02
Date of Report2019-12-31
Date of Event2019-12-09
Date Mfgr Received2019-12-09
Device Manufacturer Date2018-02-16
Date Added to Maude2020-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactW. RECKLING, M.D.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal950504482
Manufacturer Phone4082070700
Manufacturer G1SI-BONE, INC.
Manufacturer Street471 EL CAMINO REAL SUITE 101
Manufacturer CitySANTA CLARA CA 950504482
Manufacturer CountryUS
Manufacturer Postal Code950504482
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIFUSE IMPLANT SYSTEM
Generic NameORTHOPEDIC ROD
Product CodeOUR
Date Received2020-01-02
Model Number7065M-100
Lot Number2628271
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSI-BONE, INC.
Manufacturer Address471 EL CAMINO REAL SUITE 101 SANTA CLARA CA 950504482 US 950504482


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.