VIEWFLEX? PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO 100043720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-02 for VIEWFLEX? PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO 100043720 manufactured by St. Jude Medical, Inc..

Event Text Entries

[178026719] One viewflex? Plus ice catheter interface module for viewmate z intracardiac ultrasound console was received for investigation. Trace amounts of foreign material was identified on the transducer connector port. Disassembly was performed, and heavy amounts of saline deposits were identified on numerous internal assemblies with signs of arcing between electronic components. The device history record was reviewed to ensure that each manufacturing and inspection operation was performed. Based on the information provided to abbott and the investigation performed, there was fluid ingress to the internal assemblies resulting in hardware damage.
Patient Sequence No: 1, Text Type: N, H10


[178026720] This event is being reported based on the analysis of the returned device.?
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2184149-2019-00259
MDR Report Key9540330
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-02
Date of Report2020-01-02
Date of Event2019-09-24
Date Mfgr Received2019-12-11
Device Manufacturer Date2011-09-02
Date Added to Maude2020-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
Manufacturer StreetONE ST. JUDE MEDICAL DRIVE
Manufacturer CityST. PAUL MN 55117
Manufacturer CountryUS
Manufacturer Postal Code55117
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIEWFLEX? PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO
Generic NameULTRASONIC PULSED DOPPLER IMAGING SYSTEM
Product CodeIYN
Date Received2020-01-02
Returned To Mfg2019-10-28
Model Number100043720
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerST. JUDE MEDICAL, INC.
Manufacturer AddressONE ST. JUDE MEDICAL DRIVE ST. PAUL MN 55117 US 55117


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.