Event Text Entries

[173724479] The koala catheter is specifically designed to be as atraumatic as possible, thus such an incident is extremely rare over its 20+ year history. An internal investigation of the incident concluded that the most likely root cause is the inadvertent advancement of the introducer beyond the opening of the cervix, as this is the only material that is stiff enough to potentially cause trauma to the fetus. The distributor will provide training to the hospital to stress the point that the introducer should not advance the providers fingers placed at the cervical opening. If any additional information is received, a follow-up report will be sent.
Patient Sequence No: 1, Text Type: N, H10


[173724480] A baby had a "bruised" scar in the face after delivery and the parents sharply complained about it to the hospital.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1722684-2020-00001
MDR Report Key9541191
Report SourceDISTRIBUTOR
Date Received2020-01-02
Date of Report2020-01-02
Date Mfgr Received2019-12-11
Date Added to Maude2020-01-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KELLIE STEFANIAK
Manufacturer Street747 W 4170 S
Manufacturer CityMURRAY UT 84123
Manufacturer CountryUS
Manufacturer Postal84123
Manufacturer Phone8012688200
Manufacturer G1CLINICAL INNOVATIONS, LLC
Manufacturer Street747 W 4170 S
Manufacturer CityMURRAY UT 84123
Manufacturer CountryUS
Manufacturer Postal Code84123
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKOALA IUPC INTRAUTERINE PRESSURE CATHETER
Generic NameTRANSDUCER, PRESSURE, INTRAUTERINE
Product CodeHFN
Date Received2020-01-02
Model NumberIPC-5000E
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCLINICAL INNOVATIONS, LLC
Manufacturer Address747 W 4170 S MURRAY UT 84123 US 84123


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-02

© 2026 FDA.report