MONOBLOCK INSERTER HANDLE N/A 110024780

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2020-01-03 for MONOBLOCK INSERTER HANDLE N/A 110024780 manufactured by Zimmer Biomet, Inc..

Event Text Entries

[174412223] (b)(4). Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[174412224] It was reported the surgeon was implanting the arcos one-piece stem into the right hip, the threaded tip of the inserter hand broke off into the prosthesis about a third of the way down of being fully implanted. Surgeon proceeded to try to remove the tip from the stem but couldn't because all that was exposed in the insertion hole was a flat smooth surface. About 2 mm of the tip broke off. Since it was impossible to remove the broken instrument, the surgeon decided to leave it embedded in the implant since it was safely stuck in the stem. Attempts have been made and additional information on the reported event is unavailable at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001825034-2020-00018
MDR Report Key9544781
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2020-01-03
Date of Report2020-01-02
Date of Event2019-12-09
Date Mfgr Received2019-12-13
Date Added to Maude2020-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER BIOMET, INC.
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameMONOBLOCK INSERTER HANDLE
Generic NameORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Product CodeHXC
Date Received2020-01-03
Model NumberN/A
Catalog Number110024780
Lot Number744190
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET, INC.
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-03