THERMOPHORE 155

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-01-03 for THERMOPHORE 155 manufactured by Battle Creek Equipment.

Event Text Entries

[180777757] Customer stated, "pad will not shut off. " customer did not claim injury. Product was returned. Investigator determined through testing that the switch when put in the off position would still return to drawing power as if on the high setting. Further investigation will be conducted by switch manufacturer, along with a follow up report submitted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1832415-2019-10983
MDR Report Key9545665
Report SourceCONSUMER
Date Received2020-01-03
Date of Report2019-12-06
Date Added to Maude2020-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMICHAEL GREEN
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHERMOPHORE
Generic NameTHERMOPHORE
Product CodeIRT
Date Received2020-01-03
Returned To Mfg2019-12-16
Model Number155
Lot Number00155037193
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBATTLE CREEK EQUIPMENT
Manufacturer Address702 S REED RD FREMONT IN 46737 US 46737


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-03

© 2024 FDA.report