ALERE AFINION AS100 ANALYZER 1116053

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-01-03 for ALERE AFINION AS100 ANALYZER 1116053 manufactured by Abbott Diagnositcs Technologies As.

Event Text Entries

[188869021] A investigation is ongoing for afinion analyzers related to similar incidents and the investigation to date has concluded that electric shock has not been demonstrated with the analyzers. Rather, static discharge may be possible and is more consistent with the description of the incident. Per the health hazard evaluation static discharge does not pose a risk to harm. Although there was no injury reported in this complaint, due to the burnt smell this is considered a potential hazard, which could have th potential to result in burn injury.
Patient Sequence No: 1, Text Type: N, H10


[188869022] The customer had just removed a test cassette from the analyzer and stretched her hand over the instrument to pick up a glove. Suddenly there was a sound from the instrument and a burnt smell. There was no smoke. The power went off in the whole office. The customer reported that she felt an electric shock in her arm and shoulder and felt a bit numb for a while. When speaking to the customer around 45 minutes after she said she felt fine. No permanent injury was reported. The customer says that she seems to be "an electrical person" since it's not the first time she has experiences similar types of incidents, but not related to afinion instrument. Previous incidents have not been as powerful as this incident. She sometimes experiences similar but small electric shock when touching other equipment. When the customer went back and touched the bench where the instrument was placed she still experienced small electrical shocks.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003045237-2019-00006
MDR Report Key9545705
Report SourceUSER FACILITY
Date Received2020-01-03
Date of Report2020-01-03
Date of Event2019-12-11
Date Mfgr Received2019-12-11
Date Added to Maude2020-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS MONICA VALLESTAD
Manufacturer StreetKJELS P.O.BOX 6863 RODEL
Manufacturer CityOSLO, 0504
Manufacturer CountryNO
Manufacturer Postal0504
Manufacturer G1ABBOTT DIAGNOSITCS TECHNOLOGIES AS
Manufacturer StreetKJELS P.O.BOX 6863 RODEL
Manufacturer CityOSLO, 0504
Manufacturer CountryNO
Manufacturer Postal Code0504
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALERE AFINION AS100 ANALYZER
Generic NameANALYZER
Product CodeJQT
Date Received2020-01-03
Model Number1116053
Catalog Number1116053
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT DIAGNOSITCS TECHNOLOGIES AS
Manufacturer AddressKJELS?SVEIEN 161 P.O.BOX 6863 RODEL?KKA OSLO, 0504 NO 0504


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-03