HELIOS H46T B-700847-00

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2020-01-03 for HELIOS H46T B-700847-00 manufactured by Caire Inc..

Event Text Entries

[188211223] The unit has been returned for an evaluation. If any new information is discovered, a follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[188211224] After the filling process at the patient's home, the unit iced up. Pangas retrieved the device for inspection in their repair centre. Their inspection revealed that the cable to the filling status display (gauge showing level of o2 in the tank) had burst and liquid oxygen had escaped from it. Because of continuous high pressure, lox flowed through the red tube which is directly connected to the display and caused the icing. Additional comments on 10th december 2019: the event happened during the filling process by the heimox operative in the technical room of a hospital. The tank was immediately taken out of service and therefore no patient was affected by the incident. There was no injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004972304-2019-00055
MDR Report Key9548784
Report SourceFOREIGN
Date Received2020-01-03
Date of Report2020-02-05
Date of Event2019-11-20
Date Mfgr Received2019-12-04
Date Added to Maude2020-01-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. NEAL MALOY
Manufacturer Street2200 AIRPORT INDUSTRIAL DRIVE SUITE 500
Manufacturer CityBALL GROUND GA 30107
Manufacturer CountryUS
Manufacturer Postal30107
Manufacturer Phone7707217700
Manufacturer G1CAIRE INC.
Manufacturer Street2200 AIRPORT INDUSTRIAL DRIVE SUITE 500
Manufacturer CityBALL GROUND GA 30107
Manufacturer CountryUS
Manufacturer Postal Code30107
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHELIOS H46T
Generic NameUNIT, LIQUID OXYGEN, STATIONARY
Product CodeBYJ
Date Received2020-01-03
Returned To Mfg2019-12-10
Model NumberB-700847-00
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAIRE INC.
Manufacturer Address2200 AIRPORT INDUSTRIAL DRIVE SUITE 500 BALL GROUND GA 30107 US 30107


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.