MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-01-06 for NEPTUNE 2 DOCKING STATION (120V) 0702014000 manufactured by Stryker Instruments-kalamazoo.
[173771666]
Per the service inspection, it is likely fluid was spilled onto the plug of the device where it mates with the wall outlet, causing the event.
Patient Sequence No: 1, Text Type: N, H10
[173771667]
It was reported that during docking the device after a surgical procedure the device short circuited and sparked. No medical intervention and no adverse consequences were reported with this event. As this event occurred after a surgical procedure, there was no patient involvement and no delay to a surgical procedure.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 0001811755-2020-00012 |
| MDR Report Key | 9553093 |
| Report Source | USER FACILITY |
| Date Received | 2020-01-06 |
| Date of Report | 2020-01-06 |
| Date of Event | 2019-12-16 |
| Date Mfgr Received | 2019-12-16 |
| Device Manufacturer Date | 2019-01-29 |
| Date Added to Maude | 2020-01-06 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Reporter Occupation | BIOMEDICAL ENGINEER |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | MR. ZACH BAKER |
| Manufacturer Street | 4100 EAST MILHAM AVENUE |
| Manufacturer City | KALAMAZOO MI 49001 |
| Manufacturer Country | US |
| Manufacturer Postal | 49001 |
| Manufacturer Phone | 2693237700 |
| Manufacturer G1 | STRYKER INSTRUMENTS-KALAMAZOO |
| Manufacturer Street | 4100 EAST MILHAM AVENUE |
| Manufacturer City | KALAMAZOO MI 49001 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 49001 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | NEPTUNE 2 DOCKING STATION (120V) |
| Generic Name | APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED |
| Product Code | JCX |
| Date Received | 2020-01-06 |
| Catalog Number | 0702014000 |
| Device Availability | Y |
| Device Eval'ed by Mfgr | Y |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | STRYKER INSTRUMENTS-KALAMAZOO |
| Manufacturer Address | 4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2020-01-06 |