Event Text Entries

[173771666] Per the service inspection, it is likely fluid was spilled onto the plug of the device where it mates with the wall outlet, causing the event.
Patient Sequence No: 1, Text Type: N, H10


[173771667] It was reported that during docking the device after a surgical procedure the device short circuited and sparked. No medical intervention and no adverse consequences were reported with this event. As this event occurred after a surgical procedure, there was no patient involvement and no delay to a surgical procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001811755-2020-00012
MDR Report Key9553093
Report SourceUSER FACILITY
Date Received2020-01-06
Date of Report2020-01-06
Date of Event2019-12-16
Date Mfgr Received2019-12-16
Device Manufacturer Date2019-01-29
Date Added to Maude2020-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. ZACH BAKER
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal49001
Manufacturer Phone2693237700
Manufacturer G1STRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Street4100 EAST MILHAM AVENUE
Manufacturer CityKALAMAZOO MI 49001
Manufacturer CountryUS
Manufacturer Postal Code49001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEPTUNE 2 DOCKING STATION (120V)
Generic NameAPPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Product CodeJCX
Date Received2020-01-06
Catalog Number0702014000
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-06

© 2024 FDA.report