ALINITY C CALCIUM REAGENT 07P5720 07P57-20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2020-01-06 for ALINITY C CALCIUM REAGENT 07P5720 07P57-20 manufactured by Abbott Gmbh.

Event Text Entries

[188236271] Complete information for section a. Patient information, 1. Patient identifier = sid685063673301 an evaluation is in process. A follow-up report will be submitted when the evaluation is complete. There is no further patient information provided due to privacy issues.
Patient Sequence No: 1, Text Type: N, H10


[188236272] The customer reported falsely elevated calcium results on one male patient generated on the alinity c analyzer. The results generated were: on (b)(6) 2019 sid685063673301 on serial number (b)(4) = 5. 8490mmol/l (normal range 2. 15-2. 50mmol/l) / retest on sn (b)(4) = 4. 8167 / sn (b)(4) = >6. 0; the sample was then poured into cups - serum on sn (b)(4) = 3. 9814mmol/l / plasma 3. 8804 / plasma on sn (b)(4)= 3. 7996mmol/l; new sample drawn = 3. 99mmol/l / tested on abl800 analyzer = 2. 25mmol/l. There was no reported impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002809144-2020-00012
MDR Report Key9554429
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2020-01-06
Date of Report2020-02-25
Date of Event2019-12-04
Date Mfgr Received2020-02-25
Device Manufacturer Date2019-02-25
Date Added to Maude2020-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCHRISTIAN LEE
Manufacturer Street100 ABBOTT PARK ROAD DEPT 09B9, BLDG CP01-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224668-294
Manufacturer G1ABBOTT GMBH
Manufacturer StreetMAX-PLANCK-RING 2
Manufacturer CityWIESBADEN 65205
Manufacturer CountryGM
Manufacturer Postal Code65205
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALINITY C CALCIUM REAGENT
Generic NameCALCIUM
Product CodeCJY
Date Received2020-01-06
Model Number07P5720
Catalog Number07P57-20
Lot Number63032UN18
Device Expiration Date2019-12-11
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT GMBH
Manufacturer AddressMAX-PLANCK-RING 2 WIESBADEN 65205 GM 65205


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-06