CADD SOLIS HPCA PUMP 2110 21-2112-0100-51

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,user facility report with the FDA on 2020-01-06 for CADD SOLIS HPCA PUMP 2110 21-2112-0100-51 manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[176118941] One cadd solis hpca pump was received with no physical damage. No issue was found in the event history log. Accuracy test was performed on the pump. The pump was previously returned due to an allegation that it "kept infusing when trying to stop", which was not verified. The delivery accuracy tests found the pump to be within the published specification of +/-6%. The pump's expulsor will be replaced as a preventive measure to bring the delivery into more nominal of a range. The reported issue could not be replicated and the cause of the issue is unknown.
Patient Sequence No: 1, Text Type: N, H10


[176118942] Information was received indicating that a smiths medical cadd solis hpca pump was over infusing. The issue was discovered during preventative maintenance and there was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3012307300-2020-00108
MDR Report Key9555027
Report SourceCONSUMER,USER FACILITY
Date Received2020-01-06
Date of Report2020-01-06
Date Mfgr Received2019-12-11
Device Manufacturer Date2012-12-28
Date Added to Maude2020-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDAVE HALVERSON
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone7633833310
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer Street6000 NATHAN LANE NORTH
Manufacturer CityMINNEAPOLIS MN 55442
Manufacturer CountryUS
Manufacturer Postal Code55442
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCADD SOLIS HPCA PUMP
Generic NamePUMP, INFUSION, PCA
Product CodeMEA
Date Received2020-01-06
Returned To Mfg2019-12-13
Model Number2110
Catalog Number21-2112-0100-51
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address6000 NATHAN LANE NORTH MINNEAPOLIS MN 55442 US 55442


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-06

© 2026 FDA.report