TRAPEZOID RX M00510880 1088

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-06 for TRAPEZOID RX M00510880 1088 manufactured by Boston Scientific Corporation.

Event Text Entries

[173849047] (b)(4). The device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[173849048] It was reported to boston scientific corporation that a trapezoid rx basket was used during a stone extraction procedure on (b)(6) 2019. According to the complainant, during the procedure, a trapezoid basket was used in an attempt to remove a stone. However, the pull wire broke during lithotripsy. A biopsy forceps was used to remove the stone from the basket, a grasper was used to remove the trapezoid basket and two stents were placed. Reportedly, the patient experienced retropneumoperitoneum post procedure due to the difficulty in completing the procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2019-06379
MDR Report Key9555721
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-06
Date of Report2020-03-17
Date of Event2019-12-09
Date Mfgr Received2020-02-20
Device Manufacturer Date2019-10-08
Date Added to Maude2020-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1BOSTON SCIENTIFIC CORPORATION
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRAPEZOID RX
Generic NameLITHOTRIPTOR, BILIARY MECHANICAL
Product CodeLQC
Date Received2020-01-06
Returned To Mfg2020-01-22
Model NumberM00510880
Catalog Number1088
Lot Number0024561769
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-06

© 2026 FDA.report