SPACELABS MEDICAL MULTIGAS ANALYZER MODULE 91518 AND ACCESSORIES FGSP1018

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional report with the FDA on 2020-01-06 for SPACELABS MEDICAL MULTIGAS ANALYZER MODULE 91518 AND ACCESSORIES FGSP1018 manufactured by Spacelabs Medical Inc..

Event Text Entries

[177824905] The complaint device was returned for evaluation. Visual evaluation found no anomalies. Device evaluation determined that the device fall time measurement failed confirming the reported malfunction. The main board was refurbished. The om board was replaced. A definitive root cause was not able to be determined; however, the most probable cause is that the device had been dropped previously during shipping. No further investigation is required. This type of event will continue to be monitored. It should be noted that this is being filed based on retrospective review.
Patient Sequence No: 1, Text Type: N, H10


[177824906] Reportedly, post repair, c02 wave pattern and numbers disappear while on patient. There was no report of patient harm. No additional event or patient information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007409280-2020-00001
MDR Report Key9555734
Report SourceCONSUMER,HEALTH PROFESSIONAL
Date Received2020-01-06
Date of Report2018-12-07
Date Mfgr Received2018-12-07
Date Added to Maude2020-01-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJENNA VARGAS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal92672
Manufacturer Phone8004495328
Manufacturer G1AVANTE HEALTH SOLUTIONS
Manufacturer Street212 AVENIDA FABRICANTE
Manufacturer CitySAN CLEMENTE CA 92672
Manufacturer CountryUS
Manufacturer Postal Code92672
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSPACELABS MEDICAL MULTIGAS ANALYZER MODULE 91518 AND ACCESSORIES
Generic NameMULTIGAS ANALYZER MODULE
Product CodeCBR
Date Received2020-01-06
Returned To Mfg2018-12-12
Model NumberFGSP1018
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSPACELABS MEDICAL INC.
Manufacturer Address5150 220TH AVE SE ISSAQUAH WA 98029 US 98029


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-06

© 2026 FDA.report