INNER SHEATH, FOR 26 FR. OUTER SHEATH A22040A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-01-07 for INNER SHEATH, FOR 26 FR. OUTER SHEATH A22040A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[178781895] The suspect medical device has not yet been returned to olympus for evaluation/investigation. Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown. However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
Patient Sequence No: 1, Text Type: N, H10


[178781896] Olympus was informed that during a therapeutic holmium laser enucleation of the prostate (holep) procedure, the ceramic insulation at the distal end of the inner sheath broke off inside the patient. It was reported that the ceramic insulation was damaged by contact with a surgical equipment and that the fracture injured the prostate, resulting in a considerable bleeding. The procedure was then changed to a therapeutic transurethral resection of the prostate (turp) procedure and completed with a similar device. No fragment remained inside the patient since it was reportedly retrieved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610773-2020-00039
MDR Report Key9557056
Report SourceCOMPANY REPRESENTATIVE,DISTRI
Date Received2020-01-07
Date of Report2020-01-27
Date of Event2019-12-11
Date Mfgr Received2020-01-24
Device Manufacturer Date2014-12-01
Date Added to Maude2020-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL WLADOW
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal22045
Manufacturer Phone4940669662
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINNER SHEATH, FOR 26 FR. OUTER SHEATH
Generic NameRESECTION SHEATHS, IRRIGATION RINGS, CYSTOSCOPE SHEATHS, HYSTEROSCOPE SHEATHS
Product CodeFJL
Date Received2020-01-07
Model NumberA22040A
Catalog NumberA22040A
Lot Number14ZW
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-07

© 2026 FDA.report