BARD-PARKER STAINLESS STEEL BLADES SIZE 11, STERILE 371211-150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-01-07 for BARD-PARKER STAINLESS STEEL BLADES SIZE 11, STERILE 371211-150 manufactured by Aspen Surgical Products, Caledonia.

Event Text Entries

[175238802] Aspen surgical received a report from the distributor indicating that a bard-parker blade broke while in use during a procedure. The incident occurred at the user facility. No sample or photographic evidence was available for evaluation. No manufacturing lot number was provided for review. Bard-parker blade broke off in the patients shoulder. The blade fell off of the blade handle on insertion and then subsequently broke as the physician was attempting to retrieve it. Physician made another incision and was able to visualize the piece and retrieve it. Broken blade was recovered from the patient and was disposed of. Patient is doing well. A review of the device history record could not be completed. The most probable root cause could have been during the stamping or grinding process. Packaging process has established controls to mitigate broken or cracked blade condition, including a "medio" blade sensor that inspects 100% of packed pouches liner level prior to aluminum foil packaging. Also, excessive force applied by end user during surgery process could also cause blade condition. The following controls are in-place to mitigate "broken blade" condition at aspen surgical las piedras site: heat treatment in-process at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test. Heat treatment quality inspections at the beginning and end of each lot are inspected for dimension integrity using a pin gauge, flatness gauge and perforation length gauge, ductility test, and hardness test. Based on this information, no additional actions required. Device not available.
Patient Sequence No: 1, Text Type: N, H10


[175238803] Aspen surgical received a report from the distributor indicating that a bard-parker blade broke during a procedure. The incident occurred at the user facility. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1836161-2020-00002
MDR Report Key9557859
Report SourceDISTRIBUTOR
Date Received2020-01-07
Date of Report2019-12-10
Date of Event2019-11-26
Date Mfgr Received2019-12-10
Date Added to Maude2020-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JORDAN HACKERT
Manufacturer Street6945 SOUTHBELT DR. S.E.
Manufacturer CityCALEDONIA MI 49316
Manufacturer CountryUS
Manufacturer Postal49316
Manufacturer Phone6165367508
Manufacturer G1SAME AS ABOVE
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBARD-PARKER STAINLESS STEEL BLADES SIZE 11, STERILE
Generic NameBARD-PARKER BLADE
Product CodeGES
Date Received2020-01-07
Model Number371211-150
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerASPEN SURGICAL PRODUCTS, CALEDONIA
Manufacturer Address6945 SOUTHBELT DR. S.E. CALEDONIA MI 49316 US 49316


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2020-01-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.