MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2020-01-07 for EMPRINT CA15L2 manufactured by Covidien Medical Products.
[173943399]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[173943400]
According to the reporter, during ablation, the antenna was completely reinserted a total of two times, repositioned twice per insertion, and one additional minor adjustment but the tip of the probe broke while in the patient body after providing 5 ablations. The physician had decided to leave the tip in the patient body with patient's consent. Computed tomography (ct) scan located the tip within the liver.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3006451981-2020-00005 |
MDR Report Key | 9558482 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2020-01-07 |
Date of Report | 2020-01-07 |
Date of Event | 2019-12-30 |
Date Mfgr Received | 2019-12-30 |
Device Manufacturer Date | 2019-09-29 |
Date Added to Maude | 2020-01-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | AMY BEEMAN |
Manufacturer Street | 161 CHESHIRE LANE, SUITE 100 |
Manufacturer City | PLYMOUTH MN 55441 |
Manufacturer Country | US |
Manufacturer Postal | 55441 |
Manufacturer Phone | 7632104064 |
Manufacturer G1 | COVIDIEN MEDICAL PRODUCTS |
Manufacturer Street | BUILDING 10- NO 789 PUXING ROA |
Manufacturer City | SHANGHAI 201114 |
Manufacturer Country | CN |
Manufacturer Postal Code | 201114 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | EMPRINT |
Generic Name | SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES |
Product Code | NEY |
Date Received | 2020-01-07 |
Model Number | CA15L2 |
Catalog Number | CA15L2 |
Lot Number | S9JG027X |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Eval'ed by Mfgr | * |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COVIDIEN MEDICAL PRODUCTS |
Manufacturer Address | BUILDING 10- NO 789 PUXING ROA SHANGHAI 201114 CN 201114 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-01-07 |