URINE METER 158101310190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2020-01-07 for URINE METER 158101310190 manufactured by Unomedical Zavodskaya Street 50.

Event Text Entries

[175020714] Mdr 3007966929-2020-00002 / device 1 of 1. (b)(6). Name of hospital: (b)(6) hospital. Based on the available information, this event is deemed to be a reportable malfunction. No samples were received. A photo was received and evaluated. Hair was observed in the photo. The defect was confirmed. A batch record review was performed. The batch record review identified a discrepancy which was investigated via a ncr: "hair inside product individual package?. A root cause investigation was performed. On the basis of available information, the following potential causes for the issue? Hair inside of primary pack? Were determined: noncompliance with the rules of movements and stay in classified rooms by operators during manufacturing process. Cap does not provide full cover of hairs. No additional investigation or actions were taken. The trend will be monitored. This issue will be monitored through the post market product monitoring review process. To date no additional information has been received. Should additional information become available, a follow-up report will be submitted. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[175020715] It was reported that a "hair was found inside the sterile packaging". Photographs depicting the reported complaint issue were submitted by the complainant. The product was not used, no harm reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007966929-2020-00002
MDR Report Key9560243
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2020-01-07
Date Mfgr Received2019-11-22
Date Added to Maude2020-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JEANETTE JOHNSON
Manufacturer Street7900 TRIAD CENTER DRIVE SUITE 400
Manufacturer CityGREENSBORO NC 27409
Manufacturer CountryUS
Manufacturer Postal27409
Manufacturer Phone3365424681
Manufacturer G1UNOMEDICAL ZAVODSKAYA STREET 50
Manufacturer StreetFE UNOMEDICAL LTD MINSK REGION
Manufacturer CityMINSKAYA VOBLASTS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameURINE METER
Generic NameDEVICE, URINE FLOW RATE MEASURING, NON-ELECTRICAL, DISPOSABLE
Product CodeFFG
Date Received2020-01-07
Model Number158101310190
Lot Number325156
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerUNOMEDICAL ZAVODSKAYA STREET 50
Manufacturer AddressFE UNOMEDICAL LTD MINSK REGION MINSKAYA VOBLASTS


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-07

© 2026 FDA.report