MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2020-01-07 for AERO X manufactured by Tisport, Llc.

Event Text Entries

[174285092] We reviewed the dhr for this chair, and it passed all applicable quality tests and configuration requirements, and met all specifications as ordered when it left the facility. The alleged event occured on a sidewalk that was described as being "very uneven with holes," and the event itself was described as being the result of the caster becoming stuck in a hole. An after market power adaptor was described as being installed; it is unclear if the power adapter was in use at the time of the incident, but by using a power adaptor, the speed and response of the end user leading up to the incident could have impacted the likelyhood of a user mistake and subsequent failure. We provide many warnings and guidances in the owners manual, including the following: chapter 1: warnings - section h. Terrain: your tilite wheelchair is designed for riding over firm, relatively smooth surfaces such as concrete and asphalt outdoors and indoor flooring (wood, tile, etc. ) and carpeting. Your tilite wheelchair is not designed for riding oversand, loose soil or rough terrain. Do not operate your chair in such terrain. You may damage the wheels or axles or loosen fasteners of your chair. If you ignore these warnings, you may fall, tip over or lose control of the wheelchair and seriously injure yourself or others or damage the wheelchair. Section g. Obstacles: warning: when riding outdoors, uneven pavement, sidewalk irregularities, potholes and other obstacles and road hazards can be dangerous. When riding indoors, doorway thresholds, plush carpeting and other obstacles can be dangerous. In order to avoid these risks: 1. Always be alert for such dangers. 2. Always scan the area ahead of you as you ride. If you ignore these warnings, you may fall, tip over or lose control of the wheelchair and seriously injure yourself or others or damage the wheelchair. Section i. Hills/slopes/ramps: warning: the center of balance of your chair is affected by the slope of the surface over which you are riding. Because your center of balance will be affected, your chair will be less stable when it is at an angle. This is especially true when riding on a slope sideways. Furthermore, do not assume anti-tips will prevent a fall or tip-over under these conditions. We have not been provided any addition information about the event or the condition of the customer at this time. A follow up medwatch form 3500a will be submitted if any additional information is provided.
Patient Sequence No: 1, Text Type: N, H10


[174285093] The end user's mother alleges that while the patient was on an uneven sidewalk with holes, the caster contacted and became stuck in a hole and a bolt in the fork assembly broke, causing the patient to fall forward and break her teeth.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3032618-2020-00001
MDR Report Key9560582
Report SourceOTHER,USER FACILITY
Date Received2020-01-07
Date of Report2020-01-07
Date of Event2019-12-09
Date Mfgr Received2019-12-09
Device Manufacturer Date2018-05-15
Date Added to Maude2020-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN ENGLISH
Manufacturer Street2701 W COURT ST
Manufacturer CityPASCO WA 99301
Manufacturer CountryUS
Manufacturer Postal99301
Manufacturer Phone5095866117
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAERO X
Generic NameWHEELCHAIR
Product CodeIOR
Date Received2020-01-07
Model NumberAERO X
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTISPORT, LLC
Manufacturer Address2701 W COURT ST PASCO WA 99301 US 99301


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-07