[177625598] Ra has just received the incident device and has been assigned to engineering for evaluation. A follow-up report will be sent upon completion of investigation.
Patient Sequence No: 1, Text Type: N, H10


[177625599] Complaint 1 of 2: (b)(4) - clip 1. Complaint 2 of 2: (b)(4) - clip 2. Name of procedure: heart procedure. Limited information available at the time of the report. In the middle of the procedure they were trying to place 2 stealth spring clips and they would not stay closed. These clips were removed. They pulled 2 more clips from another lot number, these did stay closed and case completed. There was no damage to the vessel. There was no loss of occlusion due to the clips not staying closed. The clips are available for return. Intervention: replaced with new clip of a different lot number. Patient status: no patient injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027111-2020-00303
MDR Report Key9561719
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-07
Date of Report2020-02-24
Date of Event2019-12-18
Date Mfgr Received2019-12-18
Device Manufacturer Date2019-06-04
Date Added to Maude2020-01-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactWENDY KOBAYASHI
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone9497138059
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameA1603, 6MM L/L STEALTH CLIP 3/4 F 10/BX
Generic NameCLIP, VASCULAR
Product CodeDSS
Date Received2020-01-07
Returned To Mfg2020-01-02
Model NumberA1603
Catalog Number101374101
Lot Number1357103
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAPPLIED MEDICAL RESOURCES
Manufacturer Address22872 AVENIDA EMPRESA RANCHO SANTA MARGARITA CA 92688 US 92688


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-07

© 2026 FDA.report