BALLARD CLOSED SUCTION SYSTEM FOR NEONATES/PEDIATRICS, 8 F, Y-ADAPTER 198 109382603

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-09 for BALLARD CLOSED SUCTION SYSTEM FOR NEONATES/PEDIATRICS, 8 F, Y-ADAPTER 198 109382603 manufactured by Avanos Medical Inc..

Event Text Entries

[178729903] The device history record for the reported lot number, m19120t502, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements. The product involved in the report has been returned and is being processed for evaluation. All information reasonably known as of 08-jan-2020 has been included in this health authority report. Should additional information be obtained, a follow-up health authority report will be provided. Avanos medical, inc. Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred. This product incident is documented in the avanos medical, inc. Complaint database and identified as complaint (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[178729904] It was reported that when the closed suction catheter was turned on it was not "closing off the suction to the system. " the device was had to be replaced and was in use for less than 48 hours. There was no reported injury. Additional information received 16-dec-2019 stated this incident occurred on late on (b)(6) 2019 or the early hours of the morning on (b)(6) 2019. Since the patient did not have any chest rise, she was taken off the ventilator several times to be manually bagged until the issue was properly identified. Once the issue of the failure was identified the device was changed out and the patient was able to go back on the ventilator without any issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8030647-2020-00002
MDR Report Key9568496
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-09
Date of Report2020-03-20
Date of Event2019-12-09
Date Mfgr Received2020-02-24
Device Manufacturer Date2019-04-30
Date Added to Maude2020-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LISA CLARK
Manufacturer Street5405 WINDWARD PARKWAY
Manufacturer CityALPHARETTA GA 30004
Manufacturer CountryUS
Manufacturer Postal30004
Manufacturer Phone4704485444
Manufacturer G1AVENT S DE RL DE CV
Manufacturer StreetCARRETERA INTERNACIONAL SALIDA NORTE NO. 1053
Manufacturer CityMAGDALENA, CP
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBALLARD CLOSED SUCTION SYSTEM FOR NEONATES/PEDIATRICS, 8 F, Y-ADAPTER
Generic NameVAP CLOSED SUCTION CATHETERS & ACCESSORIES
Product CodeBSY
Date Received2020-01-09
Returned To Mfg2019-12-23
Model Number198
Catalog Number109382603
Lot NumberM19120T502
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerAVANOS MEDICAL INC.
Manufacturer Address5405 WINDWARD PARKWAY ALPHARETTA GA 30004 US 30004


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-09

© 2026 FDA.report