MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,distri report with the FDA on 2020-01-09 for EMPRINT CA20L2 manufactured by Covidien Medical Products.
[174784718]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[174784719]
According to the reporter, during liver ablation, the surgeon had difficulties to remove the antenna from the hole after ablating the liver. Antenna was removed from the liver but still inside the patient's cavity. The antenna was bent accidentally and broken into pieces. Investigation of the photo submitted showed the antenna was broken into pieces. Broken pieces were successfully removed from the patient's cavity using laparoscopic forceps. There was no patient injury.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3006451981-2020-00007 |
MDR Report Key | 9569744 |
Report Source | COMPANY REPRESENTATIVE,DISTRI |
Date Received | 2020-01-09 |
Date of Report | 2020-02-17 |
Date of Event | 2020-01-07 |
Date Mfgr Received | 2020-02-03 |
Device Manufacturer Date | 2019-09-04 |
Date Added to Maude | 2020-01-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | AMY BEEMAN |
Manufacturer Street | 161 CHESHIRE LANE, SUITE 100 |
Manufacturer City | PLYMOUTH MN 55441 |
Manufacturer Country | US |
Manufacturer Postal | 55441 |
Manufacturer Phone | 7632104064 |
Manufacturer G1 | COVIDIEN MEDICAL PRODUCTS |
Manufacturer Street | BUILDING 10- NO 789 PUXING ROA |
Manufacturer City | SHANGHAI 201114 |
Manufacturer Country | CN |
Manufacturer Postal Code | 201114 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | EMPRINT |
Generic Name | SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES |
Product Code | NEY |
Date Received | 2020-01-09 |
Model Number | CA20L2 |
Catalog Number | CA20L2 |
Lot Number | S9JG008X |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | COVIDIEN MEDICAL PRODUCTS |
Manufacturer Address | BUILDING 10- NO 789 PUXING ROA SHANGHAI 201114 CN 201114 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2020-01-09 |