SOMATOM FORCE 10742326

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-09 for SOMATOM FORCE 10742326 manufactured by Siemens Healthcare Gmbh.

Event Text Entries

[175005891] (b)(6). (b)(4). Siemens completed an investigation of the reported event. The somatom force system is currently running software version va50a. The scan mode used during the procedure has a known issue with high table pitch. This issue has been solved with updated software version vb20. The affected system will be updated to vb20 on january 28, 2020. The corresponding recall has been reported to the national competent authorities including the fda separately as ct035/19/s (fda z-2274-2019).
Patient Sequence No: 1, Text Type: N, H10


[175005892] It was reported to siemens that a (b)(6) patient was initially scanned with the somatom force ct system; however, there was an image reconstruction system (irs) error and the image could not be obtained. The technician tried again to rescan but was unsuccessful. The technician the restarted the ct system and was able to complete the examination. There was no report of patient injury; however, the patient received additional x-ray dosage due to the reported issue. Therefore, this report is being submitted with an abundance of caution. The reported event occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004977335-2020-11469
MDR Report Key9570734
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-09
Date of Report2019-12-11
Date of Event2019-12-11
Date Mfgr Received2019-12-11
Date Added to Maude2020-01-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactREBECCA TUDOR
Manufacturer Street40 LIBERTY BLVD. 65-1A
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6104486484
Manufacturer G1SIEMENS HEALTHCARE GMBH
Manufacturer StreetSIEMENSTRASSE 1
Manufacturer CityFORCHHEIM, 91301
Manufacturer CountryGM
Manufacturer Postal Code91301
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-2274-2019
Event Type3
Type of Report3

Device Details

Brand NameSOMATOM FORCE
Generic NameSYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Product CodeJAK
Date Received2020-01-09
Model Number10742326
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE GMBH
Manufacturer AddressSIEMENSTRASSE 1 FORCHHEIM, 91301 GM 91301


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.