TRACTION BOOT XLARGE 5855-865

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-01-10 for TRACTION BOOT XLARGE 5855-865 manufactured by Mizuho Orthopedic Systems, Inc..

Event Text Entries

[183883481] Inspection of the returned equipment found that a threaded stud had broken off the foot plate. The consequence of this is that the boot could separate from the traction unit. This could allow a patient's leg to fall, with a potential for a serious injury. More detailed inspection of the broken parts revealed a welding defect.
Patient Sequence No: 1, Text Type: N, H10


[183883482] The user reported that during a procedure to repair a hip fracture, while the traction was being adjusted, a loud pop was heard. The equipment was checked, and nothing looked unusual, so the procedure continued. Later, when the patient's foot was rotated, it would not hold position. When the user attempted to tighten the handle more, a screw came off the boot. No patient injury was noted. While the prior reduction could not be achieved, the procedure was completed without the traction device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2921578-2020-00002
MDR Report Key9574535
Report SourceUSER FACILITY
Date Received2020-01-10
Date of Report2020-01-10
Date of Event2019-09-17
Date Mfgr Received2019-12-23
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE LEBLANC
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal945871234
Manufacturer Phone5104291500
Manufacturer G1MIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Street30031 AHERN AVENUE
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal Code945871234
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRACTION BOOT XLARGE
Generic NameBOOT
Product CodeJEA
Date Received2020-01-10
Model Number5855-865
Catalog Number5855-865
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Address30031 AHERN AVENUE UNION CITY CA 945871234 US 945871234


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-10