MAGNUS CARB.-FIBRE TABLE TOP, LONG 118016A2 - MAGNUS CARB.-FIBRE TABLE TOP, LONG BOLU 118016A2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-10 for MAGNUS CARB.-FIBRE TABLE TOP, LONG 118016A2 - MAGNUS CARB.-FIBRE TABLE TOP, LONG BOLU 118016A2 manufactured by Maquet Gmbh.

Event Text Entries

[174942767] It was reported that the carbon table top was damaged. A getinge? Maquet service technician was on site and inspected the product in question. It was detected that edges of the carbon table top were chipped off. This kind of damage is caused due to collision of the table top and obstacles under the table top or when colliding with obstacles against the table top. In the instructions for use (ifu) the user is warned concerning the risks related to collisions and advised as follows: "during the adjustment procedures, always pay attention to the or table and accessories and avoid collisions. Ensure that tubes, cables and drapes are not trapped. " besides the user is told in the ifu not to use a defective product. We assume that this issue occurred due to a use error. The product was damaged due to a use error and afterwards the product was used despite the mentioned damages. No serious injury or required medical intervention was reported. Since the extent of the reported cut is unknown, we err on the side of caution and report this as serious injury. Getinge-maquet gmbh provides product failure investigation, analysis and resolution for the device described in this report. The complete event site name is queen elizabeth university hospital glasgow and the postal code is g51 4tf.
Patient Sequence No: 1, Text Type: N, H10


[174942768] It was reported that the table top was damaged due to a collision. A user has cut his or her hand on the damaged surface during cleaning. Manufacturer reference# (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010652-2020-00002
MDR Report Key9574991
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-10
Date of Report2020-01-10
Date of Event2020-01-06
Date Mfgr Received2020-01-06
Device Manufacturer Date2014-04-15
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET GMBH KEHLER STRASSE 31,
Manufacturer CityRASTATT 76437
Manufacturer CountryGM
Manufacturer Postal76437
Manufacturer G1HOLGER ULLRICH
Manufacturer StreetMAQUET GMBH KEHLER STRASSE 31,
Manufacturer CityRASTATT 76437
Manufacturer CountryGM
Manufacturer Postal Code76437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAGNUS CARB.-FIBRE TABLE TOP, LONG
Generic NameTABLE, RADIOLOGIC
Product CodeKXJ
Date Received2020-01-10
Model Number118016A2 - MAGNUS CARB.-FIBRE TABLE TOP, LONG BOLU
Catalog Number118016A2
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET GMBH
Manufacturer AddressKEHLER STRASSE 31 RASTATT US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-10