ACCELERATOR A3600 ACP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2020-01-10 for ACCELERATOR A3600 ACP manufactured by Inpeco Sa.

Event Text Entries

[183327154] A sample tube got stuck inside the robot gripper of the rack input module (rim) which is an automation system module which allows the movement of sample tubes from a rack to the automation sample carriers. When the gripper tried to place the tube into the carrier the error on the up-down robot movement was generated and the module stopped working. The error recovery required the intervention on the module and the customer contacted the service assistance. The involved tube was a square base tube which is not allowed on the automation system: the tube was erroneously placed into the rack loaded in the rim by the operator. The field service engineer removed the safety cover and tried to pull the tube out of the gripper. Doing so, he cut his finger with the gripper fingers. He was wearing ppe, including gloves. The cut did not require any medical treatment. A blood test was performed the day after the incident: all test results were negative. The blood tests will be repeated after 1 month, 3 months, 6 months and 12 months.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3010825766-2020-00001
MDR Report Key9575273
Report SourceDISTRIBUTOR
Date Received2020-01-10
Date of Report2020-01-10
Date of Event2019-12-04
Date Mfgr Received2019-12-13
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS EVA BALZAROTTI
Manufacturer StreetVIA TORRACCIA 26
Manufacturer CityNOVAZZANO, 6883
Manufacturer CountrySZ
Manufacturer Postal6883
Manufacturer G1INPECO SPA
Manufacturer StreetVIA GIVOLETTO 15
Manufacturer CityVAL DELLA TORRE, 10040
Manufacturer CountryIT
Manufacturer Postal Code10040
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACCELERATOR A3600
Generic NameLABORATORY AUTOMATION SYSTEM
Product CodeCEM
Date Received2020-01-10
Model NumberACP
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerINPECO SA
Manufacturer AddressVIA TORRACCIA 26 NOVAZZANO, 6883 SZ 6883


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.