ULTRASOUND TRANSDUCER FOR USE WITH FM20/FM30 AVALON M2736A 862199

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2020-01-10 for ULTRASOUND TRANSDUCER FOR USE WITH FM20/FM30 AVALON M2736A 862199 manufactured by Philips Medical Systems.

Event Text Entries

[174656258] A follow up report will be submitted once the investigation is complete. Date of incident has been requested not available at time of report.
Patient Sequence No: 1, Text Type: N, H10


[174656259] The customer reported that the report outlines that a fetal heart recording of 192 beats per minute (bpm) was provided by the fetal monitor fm30 and a decision was made to perform a caesarean section. However, upon auscultation it was found to be 80 bpm. A decision was made to perform a caesarean section. It is unknown, if there was any harm to the patient or user.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610816-2020-00015
MDR Report Key9575291
Report SourceFOREIGN,USER FACILITY
Date Received2020-01-10
Date of Report2020-01-03
Date of Event2019-10-02
Date Mfgr Received2020-01-03
Device Manufacturer Date2013-03-14
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. BETTY HARRIS
Manufacturer StreetHEWLETT-PACKARD STR.2
Manufacturer CityBOEBLINGEN 71034
Manufacturer CountryGM
Manufacturer Postal71034
Manufacturer G1PHILIPS MEDICAL SYSTEMS
Manufacturer Street3000 MINUTEMAN ROAD
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal Code01810
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameULTRASOUND TRANSDUCER FOR USE WITH FM20/FM30 AVALON
Generic NameULTRASOUND TRANSDUCER
Product CodeHGM
Date Received2020-01-10
Model NumberM2736A
Catalog Number862199
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS
Manufacturer AddressHEWLETT-PACKARD STR.2 BOEBLINGEN 71034 GM 71034


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2020-01-10