PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER PX260

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-10 for PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER PX260 manufactured by Edwards Lifesciences Dr.

Event Text Entries

[188652301] The device evaluation is anticipated. However, the complaint cannot not be confirmed without the completion of a product evaluation. A supplemental report will be forthcoming with the evaluation results when received. A review of the manufacturing records indicated that the product met specifications upon release.
Patient Sequence No: 1, Text Type: N, H10


[188652302] It was reported that when this dpt (disposable pressure transducer) was connected to a right femoral line of a (b)(6) female patient, the systolic and diastolic pressure values did not match with the pressure values obtained with a non-invasive bp cuff. The sensor was zeroed and calibrated, and the line was flushed. However, the values shown on the monitor were too high and were not consistent with the patient's condition. The line was flushed several more times, but the same problem was still present, so they requested another transducer with which the values were different and more accurate to the patient's condition. The device was available for evaluation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2015691-2020-10125
MDR Report Key9576299
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-10
Date of Report2019-12-13
Date of Event2019-12-12
Date Mfgr Received2019-12-24
Device Manufacturer Date2019-09-28
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS SAMANTHA EVELEIGH
Manufacturer Street1 EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9492503939
Manufacturer G1EDWARDS LIFESCIENCES DR
Manufacturer StreetPARQUE INDUSTRIAL DE ITABO CARR. SANCHEZ KM 18.5
Manufacturer CityHAINA, SAN CRISTOBAL
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Generic NameTRANSDUCER, PRESSURE, CATHETER TIP
Product CodeDXO
Date Received2020-01-10
Returned To Mfg2020-02-04
Model NumberPX260
Catalog NumberPX260
Lot Number62625588
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES DR
Manufacturer AddressPARQUE INDUSTRIAL DE ITABO CARR. SANCHEZ KM 18.5 HAINA, SAN CRISTOBAL DR


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.