CARELINK SMARTSYNC BASE 24970A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-10 for CARELINK SMARTSYNC BASE 24970A manufactured by Plexus Manufacturing Sdn. Bhd.

Event Text Entries

[175689009] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175689010] It was reported that the user was unable to obtain right ventricular (rv) lead or right atrial (ra) lead impedance from the programmer when initially connected to the leads during implant. It was noted that the programmer would not provide an impedance measurement when the lead was connected to the alligator clips. The user noted that when changed to vvi or aai mode, whether pacing or not, the programmer would not give a measurement or update the impedance. The user noted that when they went to the threshold screen and then exited the screen, the impedance measurement would be taken or updated. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[186213516] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[188582975] Product analysis:analysis confirmed that the impedance issue is related to a known formal investigation. Impedance value does not refresh after clicking the impedance button. The impedance measurement requires both atrial and ventricular events. Depending on the events mode, it could take a few seconds to show the measured value. If the user continues to tap the field faster than that, then the update will be delayed as every tap forces the device to take a new measurement. This issue has been resolved with a new software version. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004593495-2020-00022
MDR Report Key9577294
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2020-01-10
Date of Report2020-02-07
Date of Event2019-12-13
Date Mfgr Received2020-02-04
Device Manufacturer Date2018-12-07
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-01-10
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.