VIEWMATE? Z ULTRASOUND CONSOLE WBATTERY ZS3 H701336

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-10 for VIEWMATE? Z ULTRASOUND CONSOLE WBATTERY ZS3 H701336 manufactured by Zonare Medical Systems, Inc.

Event Text Entries

[174702810] The results, method and conclusion codes along with investigation results will be provided in the final report.
Patient Sequence No: 1, Text Type: N, H10


[174702811] During a procedure, when the viewmate system was started, the console keys were not working as the patient information could not be entered. When the catheter was connected, it would not register. The system was power cycled, the probes were replaced, and the battery was re-inserted, but the issue remained. The procedure was ultimately continued by switching to tee imaging, but this transition required increased sedation time for the patient. There were no adverse consequences to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004189859-2020-00001
MDR Report Key9577319
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-10
Date of Report2020-01-21
Date of Event2019-12-13
Date Mfgr Received2020-01-16
Device Manufacturer Date2015-12-22
Date Added to Maude2020-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEPHANIE O' SULLIVAN
Manufacturer Street5050 NATHAN LANE NORTH
Manufacturer CityPLYMOUTH MN 55442
Manufacturer CountryUS
Manufacturer Postal55442
Manufacturer Phone6517565400
Manufacturer G1ZONARE MEDICAL SYSTEMS, INC
Manufacturer Street420 NORTH BERNARDO AVENUE
Manufacturer CityMOUNTAIN VIEW CA 94043
Manufacturer CountryUS
Manufacturer Postal Code94043
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIEWMATE? Z ULTRASOUND CONSOLE WBATTERY ZS3
Generic NameSYSTEM, SIGNAL ISOLATION
Product CodeDRJ
Date Received2020-01-10
Returned To Mfg2020-01-06
Model NumberH701336
Catalog NumberH701336
Lot Number5333597
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZONARE MEDICAL SYSTEMS, INC
Manufacturer Address420 NORTH BERNARDO AVENUE MOUNTAIN VIEW CA 94043 US 94043


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.