CONTOUR NEXT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2020-01-13 for CONTOUR NEXT manufactured by Ascensia Diabetes Care Us Inc..

Event Text Entries

[174912071] The patient/family was the initial reporter, so personal information was not entered. Weight was left blank as the customer's weight was not provided. The model # provided by the customer for the affected control solution was inaccurate, therefore, this information was not captured.
Patient Sequence No: 1, Text Type: N, H10


[174912072] The customer called ascensia customer service to seek help with his contour next ez meter. During the call, three control tests were run by the customer, which gave readings of 180 mg/dl, 132 mg/dl and 146 mg/dl. The meter did not automatically mark the reading of 180 mg/dl as a control test, and so the reading will be displayed as a blood result when accessing the meter's memory. There was no allegation of an adverse event. The customer was advised to return the control solution for evaluation. Replacement meter kit and test strips were sent to the customer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1810909-2020-00022
MDR Report Key9581891
Report SourceCONSUMER
Date Received2020-01-13
Date of Report2019-12-17
Date of Event2019-12-17
Date Mfgr Received2020-03-18
Device Manufacturer Date2018-12-01
Date Added to Maude2020-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactSHWETA GULATI
Manufacturer Street100 SUMMIT LAKE DRIVE
Manufacturer CityVALHALLA, NY NY 10595
Manufacturer CountryUS
Manufacturer Postal10595
Manufacturer Phone9142361830
Manufacturer G1BIONOSTICS, INC. BUSINESS TRADE NAME: RNA MEDICAL
Manufacturer StreetREGISTRATION NUMBER: 1220649 7 JACKSON RD.
Manufacturer CityDEVENS MA 01434
Manufacturer CountryUS
Manufacturer Postal Code01434
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONTOUR NEXT
Generic NameQUALITY CONTROL MATERIAL
Product CodeJJX
Date Received2020-01-13
Returned To Mfg2020-03-18
Lot Number8BV2G35
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerASCENSIA DIABETES CARE US INC.
Manufacturer Address100 SUMMIT LAKE DRIVE VALHALLA NY 10595 US 10595


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.