SURGICAL SPRING CLIPS-DOUBLE SOFT, 6MM G-6000 100374701

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2007-04-25 for SURGICAL SPRING CLIPS-DOUBLE SOFT, 6MM G-6000 100374701 manufactured by Applied Medical Resources.

Event Text Entries

[19552639] "during coronary artery by-pass grafting g-6000 applied with applier and spring clip broke at the round hub where applier is placed - hub and spring came apart - all pieces retrieved from pt. "
Patient Sequence No: 1, Text Type: D, B5


[19615733] Evaluation: pieces from four or five different spring clip units were returned and visually inspected. The body female component, is the only broken component returned. Each of the returned body female components split in a different way; but most are split below where the body female snap with the female jaw. Conclusion/corrective action: the cause of the incident was investigated but cannot be determined without additional info. Engineering was unable to duplicate the reported incident using a fully actuated model a3218 clip applicator. The type of clip applicator used, the amount of force applied and other factors are needed to understand the condition the clip was in.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027111-2007-00023
MDR Report Key958321
Report Source05
Date Received2007-04-25
Date of Event2007-02-26
Date Facility Aware2007-03-13
Date Reported to Mfgr2007-03-13
Date Added to Maude2007-12-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSUSAN NICKERSON
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal92688
Manufacturer Phone9497138396
Manufacturer G1APPLIED MEDICAL RESOURCES CORP.
Manufacturer Street22872 AVENIDA EMPRESA
Manufacturer CityRANCHO SANTA MARGARITA CA 92688
Manufacturer CountryUS
Manufacturer Postal Code92688
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSURGICAL SPRING CLIPS-DOUBLE SOFT, 6MM
Generic NameSURGICAL SPRING CLIPS
Product CodeDSS
Date Received2007-04-25
Returned To Mfg2007-03-28
Model NumberG-6000
Catalog Number100374701
Lot Number1024039
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagY
Date RemovedU
Device Sequence No1
Device Event Key928622
ManufacturerAPPLIED MEDICAL RESOURCES
Manufacturer Address* RANCHO SANTA MARGARITA CA * US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2007-04-25