[175607780] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175689400] If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[175689401] It was reported that during an implant procedure, when the user attempted to select a hospital when programming into the patient information area, the programmer froze and after waiting five minutes would still not unfreeze. The user noted that they could see the patients rhythm, but could not select any buttons. The programmer remains in use. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[183797146] Product analysis: analysis determined that the application crashed due to the "<(>&<)>" symbol in the request. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004593495-2020-00029
MDR Report Key9583372
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-13
Date of Report2020-01-31
Date of Event2019-12-26
Date Mfgr Received2020-01-30
Device Manufacturer Date2018-12-07
Date Added to Maude2020-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactPAULA BIXBY
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635055378
Manufacturer G1PLEXUS MANUFACTURING SDN. BHD
Manufacturer StreetBAYAN LEPAS FREE INDUSTRIAL ZO
Manufacturer CityBAYAN LEPAS 11900
Manufacturer CountryMY
Manufacturer Postal Code11900
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCARELINK SMARTSYNC BASE
Generic NameANALYZER, PACEMAKER GENERATOR FUNCTION
Product CodeDTC
Date Received2020-01-13
Model Number24970A
Catalog Number24970A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPLEXUS MANUFACTURING SDN. BHD
Manufacturer AddressBAYAN LEPAS FREE INDUSTRIAL ZO BAYAN LEPAS 11900 MY 11900


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.