ASTRAL 150 - FUKUDA 27093

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2020-01-13 for ASTRAL 150 - FUKUDA 27093 manufactured by Resmed Ltd.

Event Text Entries

[188708001] The device was returned to an authorized resmed third party service center for an evaluation and service. The main circuit board was replaced to address this issue. Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to a software issue. Resmed's risk analysis for this failure mode concludes that the risk is acceptable. Resmed reference #: (b)(4). Report late due to transition from vmsr program. Establishment was unaware of fda letter dated 16 september 2019 for cbk procode status change in vmsr program until notified directly by fda on 18 december 2019 following submission of our quarterly summary report.
Patient Sequence No: 1, Text Type: N, H10


[188708002] It was reported to resmed that an astral device displayed an error message (sf189) related to a communication error. There was no patient harm or serious injury reported as a result of this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007573469-2020-00167
MDR Report Key9583666
Date Received2020-01-13
Date of Report2020-01-13
Date Facility Aware2019-07-19
Report Date2020-01-13
Date Reported to FDA2020-01-13
Date Reported to Mfgr2020-01-13
Date Mfgr Received2019-07-19
Date Added to Maude2020-01-13
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameASTRAL 150 - FUKUDA
Product CodeNOU
Date Received2020-01-13
Model Number27093
Catalog Number27093
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Age21 MO
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerRESMED LTD
Manufacturer Address1 ELIZABETH MACARTHUR DRIVE. BELLA VISTA SYDNEY, NSW 2153 AU 2153


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-13

© 2026 FDA.report