PIONEER PLUS CATHETER PPLUS120 400-0200.246

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-13 for PIONEER PLUS CATHETER PPLUS120 400-0200.246 manufactured by Philips Volcano.

Event Text Entries

[175183689] Internal reference: (b)(4). This case was reviewed and investigated according to the manufacturer? S policy. Additional information obtained indicates the physician was attempting to cross a non-tortuous cto at the posterior tibial artery. When removing the manufacturers device resistance with the guidewire (another manufacturer''s device) was experienced. The manufacturer's device and the guidewire became stuck and were removed together as a system. No medical or surgical intervention was performed to remove the devices. All portions of the manufacturer's device were accounted for upon removal. Under fluoroscopy a portion of a device was observed to remain in the patient. It was initially thought to be a portion of the manufacturer's device and later determined to be the guidewire (another manufacturer's device). Per device analysis all portions of the manufacturer's device were intact and accounted for, the device was not missing any material. The probable cause of the reported? Tip of the catheter broke off inside the patient? Could not be confirmed as no missing pieces were observed. The probable cause of the secondary issue (wire resistance) is use of a hydrophilic guidewire with the manufacturer's device, which is contraindicated. The manufacture's quick reference guide states in the system setup to use? 0. 014? Nonhydrophilic guidewires with long-coil transition.? Additionally, the quick reference guide warns:? Needle wire must be nonhydrophilic with long-coil transition.? No information available. The implant or explant dates are not applicable to this device. Not applicable for this device. The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.
Patient Sequence No: 1, Text Type: N, H10


[175183690] It was reported during a therapeutic peripheral procedure, on the second attempt to re-enter the peripheral vessel the tip of the manufactures device dislodged inside the patient. The tip is in a location that is not moving and the patient is in no danger, the tip is being left in the vessel. Further information provided it was not a portion of the manufacturer's device that separated but another manufacturer's device and remains in situ. This report is being submitted because an adverse event of separation of another manufacturer's guidewire device occurred while the manufacture's device was in use.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2939520-2020-00001
MDR Report Key9584531
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-13
Date of Report2019-11-14
Date of Event2019-11-12
Report Date2005-01-01
Date Reported to FDA2005-01-01
Date Reported to Mfgr2005-01-10
Date Mfgr Received2019-12-16
Device Manufacturer Date2019-06-18
Date Added to Maude2020-01-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MELISSA MANGUM
Manufacturer Street2870 KILGORE ROAD
Manufacturer CityRANCHO CORDOVA CA 95670
Manufacturer CountryUS
Manufacturer Postal95670
Manufacturer Phone8584650468
Manufacturer G1LAKE REGION MEDICAL
Manufacturer Street45 LEXINGTON DRIVE
Manufacturer CityLACONIA NH 03246
Manufacturer CountryUS
Manufacturer Postal Code03246
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NamePIONEER PLUS CATHETER
Generic NameCATHETER FOR CROSSING TOTAL OCCULUSIONS
Product CodePDU
Date Received2020-01-13
Returned To Mfg2019-12-16
Model NumberPPLUS120
Catalog Number400-0200.246
Lot Number4646331
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS VOLCANO
Manufacturer Address2870 KILGORE ROAD RANCHO CORDOVA CA 95670 US 95670

Device Sequence Number: 101

Product Code---
Date Received2020-01-13
Device Sequence No101
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2020-01-13