TRACER EX TREX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2020-01-14 for TRACER EX TREX manufactured by Invamex.

Event Text Entries

[186842857] There is no product malfunction associated with the injury. The adverse event appears to have been an accident that resulted from user error. The tracer wheelchair user manuals state, "always keep hands and fingers clear of moving parts to avoid injury. Do not place hand or fingers on the underside of the seat frame rail when opening or closing the wheelchair. Do not sit or transfer into the wheelchair unless it is fully open and the seat frame rails are fully seated into the side frame h-blocks. " the hospital quality and risk manager advised that the wheelchair is owned by the hospital and is kept folded up in a closed position by the front entrance for family members to use when they need to transport a patient into or out of the hospital. No medical personnel supervise the use of this chair. Since the chair is being used in a public place by individuals who have not read the manual, the users may not be mindful of the warnings addressed within the manual. Because of this, invacare recommends that a non-folding device be installed to keep the wheelchair from being folded when left unoccupied in a public place. The hospital quality and risk manager stated that they have considered getting non-folding devices for the chairs being used at the hospital entrance; however, they have limited space, so they may instead put up a sign that instructs the users to open the chair fully before use. A replacement wheelchair was sent to the hospital, and it was suggested that they put up a sign immediately to prevent any further injuries with the other chairs in their fleet. The subject wheelchair was returned to invacare for an evaluation. It was confirmed that it is possible for an individual to pinch their fingers between the seat rail and side frame. However, the chair functioned as designed.
Patient Sequence No: 1, Text Type: N, H10


[186842858] A quality and risk manager at an outpatient hospital reported that a patient fractured the middle finger on their right hand while they were attempting to sit down in the tracer wheelchair. She indicated that the patient put their hands on the seat rails of the chair and sat down before the chair was completely opened, using their weight to open up the chair the rest of the way. She stated that x-rays were taken, which confirmed the fracture, and the patient underwent physical therapy.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616091-2020-00003
MDR Report Key9586191
Report SourceHEALTH PROFESSIONAL
Date Received2020-01-14
Date of Report2018-01-05
Date Mfgr Received2018-01-05
Device Manufacturer Date2001-05-01
Date Added to Maude2020-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44035
Manufacturer CountryUS
Manufacturer Postal44035
Manufacturer Phone8003336900
Manufacturer G1INVAMEX
Manufacturer Street102 PARQUE INDUSTRIAL MANIMEX
Manufacturer CityREYNOSA, TAMAULIPAS 88780
Manufacturer CountryMX
Manufacturer Postal Code88780
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRACER EX
Generic NameWHEELCHAIR, MECHANICAL
Product CodeIOR
Date Received2020-01-14
Returned To Mfg2018-01-25
Model NumberTREX
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINVAMEX
Manufacturer Address102 PARQUE INDUSTRIAL MANIMEX REYNOSA, TAMAULIPAS 88780 MX 88780


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-14

© 2026 FDA.report