MULTIX TOP ACSS P 475525

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2020-01-14 for MULTIX TOP ACSS P 475525 manufactured by Siemens Healthcare Gmbh.

Event Text Entries

[174896126] Siemens is conducting an investigation of this event. The concerned collimator is built out of inflammable materials according to ul standards and it will self-extinguish in case of thermal effects. This behavior was confirmed during lab testing. A supplemental report will be submitted of additional information becomes available. Internal id # (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[174896127] Siemens was informed by its service organization that the user recognized fire and smoke coming out of the collimator on the multix top acss p system. The user put the fire out with an extinguisher. Siemens is unaware of any patient involvement. There are no injuries attribute to this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004977335-2020-14247
MDR Report Key9586759
Report SourceUSER FACILITY
Date Received2020-01-14
Date of Report2020-01-06
Date of Event2020-01-06
Date Mfgr Received2020-03-16
Date Added to Maude2020-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANASTASIA SOKOLOVA
Manufacturer Street40 LIBERTY BLVD. MC 65-1A
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6104486478
Manufacturer G1SIEMENS HEALTHCARE GMBH
Manufacturer StreetSIEMENSTRASSE 1
Manufacturer CityFORCHHEIM, 91301
Manufacturer CountryGM
Manufacturer Postal Code91301
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTIX TOP ACSS P
Generic NameRADIOLOGIC TABLE.
Product CodeIZZ
Date Received2020-01-14
Model Number475525
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE GMBH
Manufacturer AddressSIEMENSTRASSE 1 FORCHHEIM, 91301 GM 91301


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-01-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.