MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2020-01-14 for AXIOS STENT AND DELIVERY SYSTEM M00553550 5355 manufactured by Boston Scientific Corporation.
[176381167]
(b)(4). Although expected, the device has not been received for analysis. Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10
[176381168]
It was reported to boston scientific corporation that a hot axios stent was to be implanted in a transduodenal position to treat a pancreatic pseudocyst during a cyst drainage procedure performed on (b)(6) 2019. According to the complainant, during the procedure, after deploying the first flange of the stent, the first flange failed to expand. The physician attempted to deploy the second flange of the stent and when the delivery system was pulled back into the duodenum, the stent ended up fully deployed in an unintended location. The stent was removed from the patient using a snare. The procedure was not completed, because the physician did not have another hot axios stent available. There were no reported patient complications as a result of this event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005099803-2019-06520 |
MDR Report Key | 9587005 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2020-01-14 |
Date of Report | 2020-03-25 |
Date of Event | 2019-12-19 |
Date Mfgr Received | 2020-03-03 |
Device Manufacturer Date | 2019-02-01 |
Date Added to Maude | 2020-01-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | CAROLE MORLEY |
Manufacturer Street | 300 BOSTON SCIENTIFIC WAY |
Manufacturer City | MARLBOROUGH MA 01752 |
Manufacturer Country | US |
Manufacturer Postal | 01752 |
Manufacturer Phone | 5086834015 |
Manufacturer G1 | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Street | BALLYBRIT BUSINESS PARK |
Manufacturer City | GALWAY |
Manufacturer Country | EI |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AXIOS STENT AND DELIVERY SYSTEM |
Generic Name | PANCREATIC STENT, COVERED, METALLIC, REMOVABLE |
Product Code | PCU |
Date Received | 2020-01-14 |
Returned To Mfg | 2020-02-03 |
Model Number | M00553550 |
Catalog Number | 5355 |
Lot Number | 0023278219 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BOSTON SCIENTIFIC CORPORATION |
Manufacturer Address | 300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2020-01-14 |