CYSTOSCOPY IRRIGATION SET 2C4040

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2020-01-14 for CYSTOSCOPY IRRIGATION SET 2C4040 manufactured by Baxter Healthcare Corporation.

Event Text Entries

[175027435] (b)(4). Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[175027436] It was reported nineteen (19) cysto/bladder irrigation sets were contaminated. It was further reported the tubing appeared to have "diffuse/random greyish-black discoloration that transverses the tubing laterally (when cut open for inspection). " there was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[179864999] Additional information was added. Five (5) devices were received for evaluation. One (1) of the five samples was received having a cut on the tubing. A visual inspection was performed using the naked eye which observed a brown/yellow discoloration between the walls of the tubing, which caused a visual effect of contamination in the tubing. The discoloration was cosmetic in nature and no particulate matter (pm) was observed in the fluid path of the tubing. Therefore, the reported condition of pm in fluid path was not verified. Functional testing was not performed for this complaint. The other samples were not received for evaluation; therefore, a device analysis could not be completed. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416980-2019-07557
MDR Report Key9587715
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2020-01-14
Date of Report2020-02-17
Date of Event2019-12-16
Date Mfgr Received2020-01-27
Date Added to Maude2020-01-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CARTAGO
Manufacturer StreetSEE H10
Manufacturer CityCARTAGO 30106
Manufacturer CountryCS
Manufacturer Postal Code30106
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCYSTOSCOPY IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2020-01-14
Returned To Mfg2020-01-22
Model NumberNA
Catalog Number2C4040
Lot NumberASKU
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION
Manufacturer AddressDEERFIELD IL


Patients

Patient NumberTreatmentOutcomeDate
10 2020-01-14

© 2024 FDA.report